pH/mV Simulator (w/ screen)
HI931001
Description
HI931001 is a portable pH/mV simulators designed specifically for calibrating pH and mV (ORP) meters.
• Battery power allows for use away from a power source.
• Simulate output to discover malfunctions.
• Universal BNC connector.
Sometimes it is difficult to recognise whether a particular malfunction is due to the meter or the electrode. By simply connecting the HI-931001 to your meter’s input socket and turning the dials, pH readings can be simulated from 0 to 14 pH in 0.01 steps.
The output signals all correspond to pH values at 25°C and your pH meter should be able to check the span of your pH meter. For the mV range, the HI-931001 can simulate output from -1000 to +1000 mV in 1 mV steps.
pH SPECIFICATIONS
| pH Range : | 0.00 to 14.00 |
|---|---|
| pH Resolution : | 0.01 |
| pH Accuracy : | ±0.01 |
| pH Temperature Compensation : | All output values simulated at 25°C (77°F) |
mV SPECIFICATIONS
| mV Range : | -1000 to 1000 |
|---|---|
| mV Resolution : | 1 |
| mV Accuracy : | ±1 |
GENERAL METER SPECIFICATIONS
| Power Supply : | 9V/ 500 hours of continuous use |
|---|---|
| Environment : | 0 to 50°C (32 to 122°F); 95% RH |
| Dimensions : | 180 x 83 x 40 mm (7.1 x 3.3 x 1.6″) |
| Weight : | 320 g (11.3 oz.) |
| Ordering Information : | Supplied with HI7858/1 BNC/BNC Coaxial Cable. |
| Warranty : | 2 years |
Order now
To make your order, please contact us at info@polyscientific.com.my or call us at +6 06 3350 690